Search FDA's QSR Interpretation:     Advanced Search

 

 

 

 

 

 

Sec. 820.198b Complaint Files.

 

 

 

(b) Each manufacturer shall review and evaluate all complaints to determine whether an investigation is necessary. When no investigation is made, the manufacturer shall maintain a record that includes the reason no investigation was made and the name of the individual responsible for the decision not to investigate.

(c) Any complaint involving the possible failure of a device, labeling, or packaging to meet any of its specifications shall be reviewed, evaluated, and investigated, unless such investigation has already been performed for a similar complaint and another investigation is not necessary.

(d) Any complaint that represents an event which must be reported to FDA under part 803 of this chapter shall be promptly reviewed, evaluated, and investigated by a designated individual(s) and shall be maintained in a separate portion of the complaint files or otherwise clearly identified. In addition to the information required by § 820.198(e), records of investigation under this paragraph shall include a determination of:

(1) Whether the device failed to meet specifications;

(2) Whether the device was being used for treatment or diagnosis; and

(3) The relationship, if any, of the device to the reported incident or adverse event.

 

Interpretation:

The MDR regulation provides a mechanism for FDA and manufacturers to identify and monitor significant adverse events involving medical devices. The goals of the regulation are to detect and correct problems in a timely manner. Although the requirements of the regulation can be enforced through legal sanctions authorized by the Federal Food Drug & Cosmetic (FFD&C) Act, FDA relies on the goodwill and cooperation of all affected groups to accomplish the objectives of the regulation.

The statutory authority for the MDR regulation is section 519(a) of the FFD&C Act as amended by the Safe Medical Devices Act (SMDA) of 1990. The SMDA requires user facilities to report:

  • device-related deaths to the FDA and the device manufacturer;
  • device-related serious injuries to the manufacturer, or to FDA if the manufacturer is not known; and
  • submit to FDA on an annual basis a summary of all reports submitted during that period.

 

 

 *** for more extensive look ---- Download our 328 page guide to Interpretation of FDA's QSR with QSIT Reference!

 

Related Topics:

 

Medical Device Reporting (MDR)

Medical Device Recalls and Corrections and Removals

 

 

 

 

 

Reference: www.FDA.gov

 

The information provided here is for "public consumption" and is open to interpretation. MedicalDeviceSchool.com is not responsible for the content or the accuracy of the information. For all official interpretation and accuracy we ask that you consult the FDA's website directly and that you consult with your legal and regulatory experts on all matters of interpretation.