|
|
|
|
|||
|
|
|||||
|
|
Sec. 820.30c Design Controls. |
|
|
||
|
(c) Design
input. Each manufacturer shall establish and maintain procedures to
ensure that the design requirements relating to a device are appropriate and
address the intended use of the device, including the needs of the user and
patient. The procedures shall include a mechanism for addressing incomplete,
ambiguous, or conflicting requirements. The design input requirements shall
be documented and shall be reviewed and approved by a designated
individual(s). The approval, including the date and signature of the
individual(s) approving the requirements, shall be documented. Interpretation:
Design input is the starting point for product
design. The requirements which form the design input establish a basis for
performing subsequent design tasks and validating the design. Therefore,
development of a solid foundation of requirements is the single most
important design control activity. Many medical device manufacturers have experience
with the adverse effects that incomplete requirements can have on the design
process. A frequent complaint of developers is that "there's never time
to do it right, but there's always time to do it over." If essential
requirements are not identified until validation, expensive redesign and
rework may be necessary before a design can be released to production. By comparison, the experience of companies that
have designed devices using clear-cut, comprehensive sets of requirements is
that rework and redesign are significantly reduced and product quality is
improved. They know that the development of requirements for a medical device
of even moderate complexity is a formidable, time-consuming task. They accept
the investment in time and resources required to develop the requirements
because they know the advantages to be gained in the long run. Unfortunately, there are a number of common
misconceptions regarding the meaning and practical application of the quality
system requirements for design input. Many seem to arise from interpreting
the requirements as a literal prescription, rather than a set of principles
to be followed. In this guidance document, the focus is on explaining the
principles and providing examples of how they may be applied in typical
situations. CONCEPT DOCUMENTS VERSUS DESIGN INPUT In some cases, the marketing staff, who maintain close contact
with customers and users, determine a need for a new product, or enhancements
to an existing product. Alternatively, the idea for a new product may evolve
out of a research or clinical activity. In any case, the result is a concept
document specifying some of the desired characteristics of the new product. Some members of the medical device community view
these marketing memoranda, or the equivalent, as the design input. However,
that is not the intent of the quality system requirements. Such concept
documents are rarely comprehensive, and should not be expected to be so.
Rather, the intent of the quality system requirements is that the product
conceptual description be elaborated, expanded, and transformed into a
complete set of design input requirements which are written to an engineering
level of detail. This is an important concept. The use of
qualitative terms in a concept document is both appropriate and practical.
This is often not the case for a document to be used as a basis for design.
Even the simplest of terms can have enormous design implications. For
example, the term "must be portable" in a concept document raises
questions in the minds of product developers about issues such as size and
weight limitations, resistance to shock and vibration, the need for
protection from moisture and corrosion, the capability of operating over a
wide temperature range, and many others. Thus, a concept document may be the
starting point for development, but it is not the design input requirement.
This is a key principle-the design input requirements are the result of the
first stage of the design control process. RESEARCH AND DEVELOPMENT. Some manufacturers have difficulty in determining where
research ends and development begins. Research activities may be undertaken
in an effort to determine new business opportunities or basic characteristics
for a new product. It may be reasonable to develop a rapid prototype to
explore the feasibility of an idea or design approach, for example, prior to
developing design input requirements. But manufacturers should avoid falling
into the trap of equating the prototype design with a finished product
design. Prototypes at this stage lack safety features and ancillary functions
necessary for a finished product, and are developed under conditions which
preclude adequate consideration of product variability due to manufacturing. RESPONSIBILITY FOR DESIGN INPUT DEVELOPMENT. Regardless of who developed the initial product concept,
product developers play a key role in developing the design input
requirements. When presented with a set of important characteristics, it is
the product developers who understand the auxiliary issues that must be
addressed, as well as the level of detail necessary to design a product.
Therefore, a second key principle is that the product developer(s) ultimately
bear responsibility for translating user and/or patient needs into a set of
requirements which can be validated prior to implementation. While this is
primarily an engineering function, the support or full participation of
production and service personnel, key suppliers, etc., may be required to
assure that the design input requirements are complete. Care must be exercised in applying this principle.
Effective development of design input requirements encompasses input from
both the product developer as well as those representing the needs of the
user, such as marketing. Terminology can be a problem. In some cases, the
product conceptual description may be expressed in medical terms. Medical
terminology is appropriate in requirements when the developers and reviewers
are familiar with the language, but it is often preferable to translate the
concepts into engineering terms at the requirements stage to minimize
miscommunication with the development staff. Another problem is incorrect assumptions. Product
developers make incorrect assumptions about user needs, and marketing
personnel make incorrect assumptions about the needs of the product
designers. Incorrect assumptions can have serious consequences that may not
be detected until late in the development process. Therefore, both product
developers and those representing the user must take responsibility for
critically examining proposed requirements, exploring stated and implied
assumptions, and uncovering problems. Some examples should clarify this point. A basic
principle is that design input requirements should specify what the design is
intended to do while carefully avoiding specific design solutions at this
stage. For example, a concept document might dictate that the product be
housed in a machined aluminum case. It would be prudent for product
developers to explore why this type of housing was specified. Perhaps there
is a valid reason-superior electrical shielding, mechanical strength, or
reduced time to market as compared to a cast housing. Or perhaps machined
aluminum was specified because a competitor's product is made that way, or
simply because the user didn't think plastic would be strong enough. Not all incorrect assumptions are made by users.
Incorrect assumptions made by product developers may be equally damaging.
Failure to understand the abuse to which a portable instrument would be
subjected might result in the selection of housing materials inadequate for
the intended use of the product. There are occasions when it may be appropriate to specify
part of the design solution in the design input requirements. For example, a
manufacturer may want to share components or manufacturing processes across a
family of products in order to realize economies of scale, or simply to help
establish a corporate identity. In the case of a product upgrade, there may
be clear consensus regarding the features to be retained. However, it is
important to realize that every such design constraint reduces implementation
flexibility and should therefore be documented and identified as a possible
conflicting requirement for subsequent resolution. SCOPE AND LEVEL OF DETAIL. Design input requirements must be comprehensive. This may be
quite difficult for manufacturers who are implementing a system of design
controls for the first time. Fortunately, the process gets easier with
practice. It may be helpful to realize that design input requirements fall
into three categories. Virtually every product will have requirements of all
three types.
What is the scope of the design input requirements
development process and how much detail must be provided? The scope is
dependent upon the complexity of a device and the risk associated with its
use. For most medical devices, numerous requirements encompassing functions,
performance, safety, and regulatory concerns are implied by the application.
These implied requirements should be explicitly stated, in engineering terms,
in the design input requirements. Determining the appropriate level of detail
requires experience. However, some general guidance is possible. The
marketing literature contains product specifications, but these are superficial.
The operator and service manuals may contain more detailed specifications and
performance limits, but these also fall short of being comprehensive. Some
insight as to what is necessary is provided by examining the requirements for
a very common external interface. For the power requirements for AC-powered
equipment, it is not sufficient to simply say that a unit shall be
AC-powered. It is better to say that the unit shall be operable from AC power
in North America, Europe, and Japan, but that is still insufficient detail to
implement or validate the design. If one considers the situation just in
North America, where the line voltage is typically 120 volts, many systems
are specified to operate over the range of 108 to 132 volts. However, to
account for the possibility of brownout, critical devices may be specified to
operate from 95 to 132 volts or even wider ranges. Based on the intended use
of the device, the manufacturer must choose appropriate performance limits. There are many cases when it is impractical to
establish every functional and performance characteristic at the design input
stage. But in most cases, the form of the requirement can be determined, and
the requirement can be stated with a to-be-determined (TBD) numerical value
or a range of possible values. This makes it possible for reviewers to assess
whether the requirements completely characterize the intended use of the
device, judge the impact of omissions, and track incomplete requirements to
ensure resolution. For complex designs, it is not uncommon for the
design input stage to consume as much as thirty percent of the total project
time. Unfortunately, some managers and developers have been trained to
measure design progress in terms of hardware built, or lines of software code
written. They fail to realize that building a solid foundation saves time
during the implementation. Part of the solution is to structure the
requirements documents and reviews such that tangible measures of progress
are provided. At the other extreme, many medical devices have
very simple requirements. For example, many new devices are simply
replacement parts for a product, or are kits of commodity items. Typically,
only the packaging and labeling distinguishes these products from existing
products. In such cases, there is no need to recreate the detailed design
input requirements of the item. It is acceptable to simply cite the
predecessor product documentation, add any new product information, and
establish the unique packaging and labeling requirements. ASSESSING DESIGN INPUT REQUIREMENTS FOR ADEQUACY. Eventually, the design input must be reviewed for adequacy.
After review and approval, the design input becomes a controlled document.
All future changes will be subject to the change control procedures, as discussed
in Section I (Design Changes). Any assessment of design input requirements boils
down to a matter of judgment. As discussed in Section E (Design Review), it
is important for the review team to be multidisciplinary and to have the
appropriate authority. A number of criteria may be employed by the review
team.
EVOLUTION OF THE DESIGN INPUT REQUIREMENTS. Large development projects often are implemented in stages.
When this occurs, the design input requirements at each stage should be
developed and reviewed following the principles set forth in this section.
Fortunately, the initial set of requirements, covering the overall product,
is by far the most difficult to develop. As the design proceeds, the output
from the early stages forms the basis for the subsequent stages, and the
information available to designers is inherently more extensive and detailed.
It is almost inevitable that verification
activities will uncover discrepancies which result in changes to the design
input requirements. There are two points to be made about this. One is that
the change control process for design input requirements must be carefully
managed. Often, a design change to correct one problem may create a new
problem which must be addressed. Throughout the development process, it is
important that any changes are documented and communicated to developers so
that the total impact of the change can be determined. The change control
process is crucial to device quality. The second point is that extensive rework of the
design input requirements suggests that the design input requirements may not
be elaborated to a suitable level of detail, or insufficient resources are
being devoted to defining and reviewing the requirements. Managers can use
this insight to improve the design control process. From a design control
perspective, the number of requirements changes made is less important than
the thoroughness of the change control process. Reference: www.FDA.gov The information provided
here is for "public consumption" and is open to interpretation. MedicalDeviceSchool.com
is not responsible for the content or the accuracy of the information. For
all official interpretation and accuracy we ask that you consult the FDA's website directly and
that you consult with your legal and regulatory experts on all matters of
interpretation. |
|
||||
|
|
|
||||
|
|
|
|
|||
|
|
|
|
|
|
|