|
|
|
|
||||
|
|
||||||
|
|
Sec. 820.30g Desgin Controls. |
|
|
|
||
|
(g) Design
validation. Each manufacturer shall establish and maintain procedures for
validating the device design. Design validation shall be performed under
defined operating conditions on initial production units, lots, or batches,
or their equivalents. Design validation shall ensure that devices conform to
defined user needs and intended uses and shall include testing of production
units under actual or simulated use conditions. Design validation shall
include software validation and risk analysis, where appropriate. The results
of the design validation, including identification of the design, method(s),
the date, and the individual(s) performing the validation, shall be
documented in the DHF. Interpretation:
Whereas verification is a detailed examination of
aspects of a design at various stages in the development, design validation
is a cumulative summation of all efforts to assure that the design will
conform with user needs and intended use(s), given expected variations in
components, materials, manufacturing processes, and the use environment. VALIDATION PLANNING. Planning for validation should begin early in the design
process. The performance characteristics that are to be assessed should be
identified, and validation methods and acceptance criteria should be
established. For complex designs, a schedule of validation activities and
organizational or individual responsibilities will facilitate maintaining
control over the process. The validation plan should be reviewed for
appropriateness, completeness, and to ensure that user needs and intended
uses are addressed. VALIDATION REVIEW.
Validation may expose deficiencies in the original assumptions concerning
user needs and intended uses. A formal review process should be used to
resolve any such deficiencies. As with verification, the perception of a
deficiency might be judged insignificant or erroneous, or a corrective action
may be required. VALIDATION METHODS. Many medical devices do not require clinical trials. However,
all devices require clinical evaluation and should be tested in the actual or
simulated use environment as a part of validation. This testing should
involve devices which are manufactured using the same methods and procedures
expected to be used for ongoing production. While testing is always a part of
validation, additional validation methods are often used in conjunction with
testing, including analysis and inspection methods, compilation of relevant
scientific literature, provision of historical evidence that similar designs
and/or materials are clinically safe, and full clinical investigations or
clinical trials. Some manufacturers have historically used their
best assembly workers or skilled lab technicians to fabricate test articles,
but this practice can obscure problems in the manufacturing process. It may
be beneficial to ask the best workers to evaluate and critique the
manufacturing process by trying it out, but pilot production should simulate
as closely as possible the actual manufacturing conditions. Validation should also address product packaging
and labeling. These components of the design may have significant human
factors implications, and may affect product performance in unexpected ways.
For example, packaging materials have been known to cause electrostatic
discharge (ESD) failures in electronic devices. If the unit under test is
delivered to the test site in the test engineer's briefcase, the packaging
problem may not become evident until after release to market. Validation should include simulation of the
expected environmental conditions, such as temperature, humidity, shock and
vibration, corrosive atmospheres, etc. For some classes of device, the
environmental stresses encountered during shipment and installation far
exceed those encountered during actual use, and should be addressed during
validation. Particular care should be taken to distinguish
among customers, users, and patients to ensure that validation addresses the
needs of all relevant parties. For a consumer device, the customer, user, and
patient may all be the same person. At the other extreme, the person who buys
the device may be different from the person who routinely uses it on patients
in a clinical setting. Hospital administrators, biomedical engineers, health
insurance underwriters, physicians, nurses, medical technicians, and patients
have distinct and sometimes competing needs with respect to a device design. VALIDATION DOCUMENTATION. Validation is a compilation of the results of all validation activities. For a complex design, the detailed results may be contained in a variety of separate documents and summarized in a validation report. Supporting information should be explicitly referenced in the validation report and either included as an appendix or available in the design history file.
|
|
|||||
|
|
|
|||||
|
Reference: www.FDA.gov The information provided here is for "public consumption" and is open to interpretation. MedicalDeviceSchool.com is not responsible for the content or the accuracy of the information. For all official interpretation and accuracy we ask that you consult the FDA's website directly and that you consult with your legal and regulatory experts on all matters of interpretation. |
|
|
||||
|
|
|
|
|
|
|
|