|
|
|
|
|||
|
|
|||||
|
|
Sec. 820.30i Desgin Controls. |
|
|
||
|
(i) Design
changes. Each manufacturer shall establish and maintain procedures for
the identification, documentation, validation or where appropriate
verification, review, and approval of design changes before their implementation.
Interpretation:
There are two principal administrative elements
involved in controlling design changes:
For a small development project, an adequate
process for managing change involves little more than documenting the design
change, performing appropriate verification and validation, and keeping
records of reviews. The main objectives are ensuring that:
For projects involving more than two persons,
coordination and communication of design changes become vitally important. In
other words, manufacturers should take steps to avoid the common situation
where, for example, Jon and Marie agree to a make a change but neglect to
inform Pat of their decision. Medical device manufacturers are usually quite
comfortable with the processes of document control and change control with
respect to managing manufacturing documents. The principles of these
processes are reviewed in the following paragraphs. Subsequently, we will
explore how these may be applied to design activities. DOCUMENT CONTROL. The
features of a manufacturing document control system typically include the
following:
CHANGE CONTROL.
Manufacturing change control is usually implemented using a set of
standardized procedures similar to the following:
APPLICATION OF DOCUMENT AND CHANGE CONTROLS TO
DESIGN. The design control system has to be
concerned with the creation and revision of documents, as well as the
management of finished documents. Additional mechanisms are required to
provide needed flexibility while preserving the integrity of design documentation.
These additional mechanisms are embodied in the procedures for review and
approval of various documents. It is important that the design change procedures
always include re-verifying and re-validating the design. Fortunately, most
design changes occur early in the design process, prior to extensive design
validation. Thus, for most design changes, a simple inspection is all that is
required. The later in the development cycle that the change occurs, the more
important the validation review becomes. There are numerous cases when
seemingly innocuous design changes made late in the design phase or following
release of the design to market have had disastrous consequences. For example, a manufacturer encountered problems in
the field with a valve sticking in a ventilator due to moisture in the
breathing circuit. The problem was resolved by slightly increasing the weight
of the disc. Since the change was minor, minimal testing was performed to
verify the change. Subsequently, when the revised valves entered production,
significant numbers of valves began failing. Investigation revealed that the
heavier disc was causing the valve cage to separate due to higher inertia.
This failure mode was more serious than the original sticking problem, and
resulted in a safety recall. |
|
||||
|
|
|
||||
|
Reference: www.FDA.gov The information provided here is for "public consumption" and is open to interpretation. MedicalDeviceSchool.com is not responsible for the content or the accuracy of the information. For all official interpretation and accuracy we ask that you consult the FDA's website directly and that you consult with your legal and regulatory experts on all matters of interpretation. |
|
|
|||
|
|
|
|
|
|
|