|
|
|
|
|||
|
|
|||||
|
|
820.75
Process validation.
(a) Where the
results of a process cannot be fully verified by subsequent inspection and
test, the process shall be validated with a high degree of assurance and
approved according to established procedures. The validation activities and
results, including the date and signature of the individual(s) approving the
validation and where appropriate the major equipment validated, shall be
documented. (b) Each
manufacturer shall establish and maintain procedures for monitoring and
control of process parameters for validated processes to ensure that the
specified requirements continue to be met. (1) Each
manufacturer shall ensure that validated processes are performed by qualified
individual(s). (2) For
validated processes, the monitoring and control methods and data, the date
performed, and, where appropriate, the individual(s) performing the process
or the major equipment used shall be documented. (c) When
changes or process deviations occur, the manufacturer shall review and
evaluate the process and perform revalidation where appropriate. These
activities shall be documented. Interpretation: Whereas verification is a detailed
examination of aspects of a design at various stages in the development,
design validation is a cumulative summation of all efforts to assure that the
design will conform with user needs and intended use(s), given expected
variations in components, materials, manufacturing processes, and the use
environment. VALIDATION PLANNING. Planning
for validation should begin early in the design process. The performance
characteristics that are to be assessed should be identified, and validation
methods and acceptance criteria should be established. For complex designs, a
schedule of validation activities and organizational or individual responsibilities
will facilitate maintaining control over the process. The validation plan
should be reviewed for appropriateness, completeness, and to ensure that user
needs and intended uses are addressed. VALIDATION REVIEW. Validation
may expose deficiencies in the original assumptions concerning user needs and
intended uses. A formal review process should be used to resolve any such
deficiencies. As with verification, the perception of a deficiency might be
judged insignificant or erroneous, or a corrective action may be required. VALIDATION METHODS. Many
medical devices do not require clinical trials. However, all devices require clinical evaluation and
should be tested in the actual or simulated use environment as a part of
validation. This testing should involve devices which are manufactured using
the same methods and procedures expected to be used for ongoing production.
While testing is always a part of validation, additional validation methods
are often used in conjunction with testing, including analysis and inspection
methods, compilation of relevant scientific literature, provision of
historical evidence that similar designs and/or materials are clinically
safe, and full clinical investigations or clinical trials. Some manufacturers have historically used their best assembly
workers or skilled lab technicians to fabricate test articles, but this
practice can obscure problems in the manufacturing process. It may be
beneficial to ask the best workers to evaluate and critique the manufacturing
process by trying it out, but pilot production should simulate as closely as
possible the actual manufacturing conditions. Validation should also address product packaging and labeling.
These components of the design may have significant human factors implications,
and may affect product performance in unexpected ways. For example, packaging
materials have been known to cause electrostatic discharge (ESD) failures in
electronic devices. If the unit under test is delivered to the test site in
the test engineer’s briefcase, the packaging problem may not become evident
until after release to market. Validation should include simulation of the expected
environmental conditions, such as temperature, humidity, shock and vibration,
corrosive atmospheres, etc. For some classes of device, the environmental
stresses encountered during shipment and installation far exceed those
encountered during actual use, and should be addressed during validation. Particular care should be taken to distinguish among customers,
users, and patients to ensure that validation addresses the needs of all
relevant parties. For a consumer device, the customer, user, and patient may
all be the same person. At the other extreme, the person who buys the device
may be different from the person who routinely uses it on patients in a
clinical setting. Hospital administrators, biomedical engineers, health
insurance underwriters, physicians, nurses, medical technicians, and patients
have distinct and sometimes competing needs with respect to a device design. VALIDATION DOCUMENTATION. Validation
is a compilation of the results of all validation activities. For a complex
design, the detailed results may be contained in a variety of separate
documents and summarized in a validation report. Supporting information
should be explicitly referenced in the validation report and either included
as an appendix or available in the design history file.
|
|
|
||
|
|
|
||||
|
Reference: www.FDA.gov The information provided here is for "public consumption" and is open to interpretation. MedicalDeviceSchool.com is not responsible for the content or the accuracy of the information. For all official interpretation and accuracy we ask that you consult the FDA's website directly and that you consult with your legal and regulatory experts on all matters of interpretation. |
|
|
|||
|
|
|
|
|
|
|