Search FDA's QSR Interpretation:     Advanced Search

 *** for more extensive look ---- Download our 328 page guide to Interpretation of FDA's QSR with QSIT Reference!

 

 

 

 

 

 

820.86 Acceptance status.

 

Each manufacturer shall identify by suitable means the acceptance status of product, to indicate the conformance or nonconformance of product with acceptance criteria. The identification of acceptance status shall be maintained throughout manufacturing, packaging, labeling, installation, and servicing of the product to ensure that only product which has passed the required acceptance activities is distributed, used, or installed.

Interpretation:

Written instructions are necessary to assure that components, manufacturing materials, etc. are properly identified, processed, and stored when received. Written inspection and test procedures are necessary to prescribe the:

  • acceptance activities performed;
  • dates acceptance activities are performed;
  • the results;
  • signature of the individual(s) conducting acceptance activities; and
  • where appropriate, the equipment used.

Before acceptance, all components should be either physically separated (quarantined) or clearly identified as not yet accepted. The decision to separate or tag not-yet-accepted product should be made based on the characteristics of the device, the potential for mixups, plant conditions, and manufacturing practices.

Although 820.80 requires a written procedure for accepting components, the Quality System regulation in 820.5 allows discretion in the quality system. Thus a very small manufacturer, usually 10 or fewer employees, may only need very brief written acceptance procedures referencing the purchase orders and receiving tickets. As the size of the operation, the numbers of activities, and number of people involved increase, the need for comprehensive written instructions generally increases.

Acceptance Criteria

Manufacturers should have specific acceptance criteria for components. Acceptance criteria are the attributes of a component that determine its acceptability, such as appearance, dimension, purity, performance characteristics, etc. Typically, acceptance criteria are made a part of the inspection/test procedure. For example, if component specifications or a drawing adequately describe the attributes needed in order for the component to perform in its intended manner, these may be used as the acceptance criteria. If components or the suppliers of the components have a history of good performance, the components may be accepted for use after a visual check to assure they are the items intended and that they are not damaged or contaminated. Components, which need only a visual inspection, may be accepted using the purchase order data as acceptance criteria. The purchase order and/or receiving ticket should at a minimum contain the following information:

  • name of supplier;
  • description of the component or other product; and
  • quantity shipped.

For a standard component, the catalog number may be used as a description. QA personnel should determine whether the use of any "abbreviated" criteria are adequate during their audit of production rework, history records, complaint files, and service records.

Testing and Inspection of Product

The minimum acceptance activity per current practice requires that all incoming components and other product receive at least a visual inspection for contamination and/or damage and be identified as the component specified on the purchase order. A manufacturer accepting the product has the discretion to determine when and where product should be inspected, sampled, and tested for conformance to specifications depending upon the risk that failure of that component may pose. As appropriate, product may be tested and/or inspected by:

  • the supplier;
  • when received;
  • during manufacture of the device; or
  • as part of the finished device.

If components are tested as part of the finished device, the testing should be able to reveal failed and "out-of-spec" components and not just that the finished device does not meet specifications. This determination, of course, may be performed after removing the component from the device. The rejection shall be documented [820.80(b)].

Manufacturers who decide not to sample or test specific components should be able to justify that decision based on such factors as knowledge of the supplier's previous performance in providing high quality components, the component performance history, and application of the components in the device. Manufacturers may rely on component suppliers to conduct testing if the manufacturer specifies or is knowledgeable about the supplier's quality system, particularly the inspection and test programs and the supplier has specifications that properly define the manufacturer's acceptable limits for the component or material parameters. These specifications may be used to meet the device master record requirements for component specifications, if these accurately reflect the parameters, composition, and configuration required for the component to perform the function for which it was selected. Supplier specifications are usually adequate for standard components. However, a manufacturer who relies on supplier specifications usually has no control over changes in these and, therefore, should assure at an appropriate point in the manufacturing process that the components received meet the desired specifications.

If components are tested by the supplier, acceptance of components can be based on certification and review of test data submitted by the supplier for the specific components provided. Certification should accompany each lot of components. When certification is used, the manufacturer should periodically verify the validity of the certification through an assessment of the supplier.

Where historical data shows that certain components or other product have been substandard and resulted in a device failing to meet specifications, or where performance history has not been established, specific steps should be taken to assure components meet specifications. Typically, this task is accomplished by sampling and testing each lot of components to assure that the components meet specifications. Where appropriate, all significant or high risk components should be sampled and tested.

Manufacturers may test entire assemblies of components rather than individual components. If, however, testing an assembly cannot assure fitness-for-use of the components, then components should be tested on an individual or lot basis, whichever is appropriate. For example, assemblies with an internal feedback circuit could have a very marginal component. Because of the circuit design, the condition of the marginal component might not be detected by testing the entire assembly. Therefore, the feedback loop in the assembly should be opened during one of the tests, or the individual components should be tested.

When using a contract laboratory to test production components, the laboratory becomes an extension of the device manufacturer's quality system. The device manufacturer is responsible for assuring that the contractor's test and inspection procedures are acceptable. This assurance maybe obtained by audits of the laboratory, by the lab staff, and by the finished device manufacturer.

Inspection and testing will not improve the quality of components or other product; however, if the inspection and testing is appropriate and performed adequately, these activities can be used to prevent or significantly reduce the use of low-quality or defective product. Through feedback into the overall quality system, data on products will help identify basic causes of problems and lead to solutions (820.100). If problems are found, actions such as design changes, tighter acceptance criteria, supplier assessments, or change of suppliers may be appropriate.

Acceptance and Rejection Records

Adequate records shall be maintained to provide objective evidence that components were inspected and accepted, or rejected. These records are a part of the device history record and should be maintained in a format that facilitates review. The records, however, are not required to be maintained in a single file with other production history records, and are typically filed in the receiving or quality control area according to part number or component nomenclature. Small manufacturers may use purchase orders or packing slips to record acceptance and rejection if they contain adequate information.

The Quality System regulation specifies in 820.80(b) that a record of component acceptance and rejection be maintained. Typically, acceptance/rejection records should contain:

  • acceptance or rejection documentation;
  • number and type of deficiencies;
  • quantity approved;
  • quantity rejected; and
  • nature of corrective action taken.

Obsolete, Deteriorated, and Rejected Components

Obsolete, deteriorated and rejected components shall be identified (820.60, 820.86, and 820.150) as such and be placed in a separate quarantine area or specially identified area to prevent mixups. If practical, components should be individually identified as rejects. Where it is not feasible to tag each rejected component, as in the case of transistors, bolts, bottles, etc., containers or packages of rejected lots should be clearly marked and otherwise appropriately segregated from accepted components. See 820.86 for clarification. Manufacturers should determine the need for a separate written procedure for handling these components based on the size of the manufacturer and complexity of their devices and operations. Disposition of nonconforming product shall be documented [820.90(b)].

Records for rejected components should state whether the components were returned, scraped, reworked, etc. In very small manufacturers, disposition can be recorded directly onto the purchase order, receiving ticket, or other associated document. Small-to-medium sized manufacturers generally record disposition on the form used to receive components. Most large manufacturers record disposition of rejected components on standard forms such as a Nonconforming Material Report (NMR).

When components, materials, etc., become obsolete, many manufacturers assign new identification numbers to the new version of these components etc. The obsolete items are retained for other uses, such as repair parts, engineering projects, etc. In these cases, the old and new items should be adequately segregated and/or identified to prevent inadvertent use of obsolete components in production.

 

 

 

 

 

 

 

 

Reference: www.FDA.gov

 

The information provided here is for "public consumption" and is open to interpretation. MedicalDeviceSchool.com is not responsible for the content or the accuracy of the information. For all official interpretation and accuracy we ask that you consult the FDA's website directly and that you consult with your legal and regulatory experts on all matters of interpretation.