Biological Indicator Reduced Incubation Time (RIT) Study

 

cont'd

 
   

 

 

 

The greatest number of days of incubation required to obtain more than 97% positive BIs (based on the 7 day incubation time) in any one of the partial cycles is the minimum incubation time that will be allowed. Averaging the three (or more) partial cycle incubation times is not allowable (see example in Appendix).

If the biological indicator user has not or cannot validate the biological indicator incubation period using the described methodology, then the user should remain with at least a seven day incubation period.

Number Positive Biological Indicators Required to Achieve 97% Growth

 

Numerator Data1

30

31

32

33

33

34

35

36

37

38

 

Denominator Data2

30

31

32

33

34

35

36

37

38

39

 

Numerator Data

39

40

41

42

43

44

45

46

47

48

 

Denominator Data

40

41

42

43

44

45

46

47

48

49

 

Numerator Data

49

50

51

52

53

54

55

56

57

58

 

Denominator Data

50

51

52

53

54

55

56

57

58

59

 

Numerator Data

59

60

61

62

63

64

65

65

66

67

 

Denominator Data

60

61

62

63

64

65

66

67

68

69

 

Numerator Data

68

69

70

71

72

73

74

75

76

77

78

Denominator Data

70

71

72

73

74

75

76

77

78

79

80

 Start of Article                                    ...Next (more)........

 

   