Biological Indicator Reduced Incubation Time (RIT) Study

 

cont'd

 
   

 

 

 

1 The numerator is the number of positive biological indicators that is greater than 97% of the denominator. If the numerator is equal to or greater than the one listed for the corresponding denominator (based on the total number of positive biological indicators on day 7), the length of the incubation time when this occurs is acceptable.

2 The denominator is the total number of positive biological indicators on day 7 of incubation.

Example: A manufacturer would like to reduce its biological indicator incubation time to 3 days. The test data show the following:

 

Day 1

Day 2

Day 3

Day 4

Day 5

Day 6

Day 7

Partial cycle #1

 

 

 

 

 

 

 

Numerator

0

56

57

57

58

59

59

Denominator

59

59

59

59

59

59

59

Percent Growth

-

94.9

96.6

96.6

98.3

100

100

 

 

 

 

 

 

 

 

Partial cycle #2

1

34

35

35

35

35

35

Numerator

35

35

35

35

35

35

35

Denominator

2.9

97.1

100

100

100

100

100

Percent Growth

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Partial cycle #3

 

 

 

 

 

 

 

Numerator

0

79

81

Test invalid because percent positive biological indicators is outside the allowable window (30-80%

Denominator

 

 

 

Percent Growth

 

 

 

 

 

 

 

 

 

 

 

Partial cycle #4

 

 

 

 

 

 

 

Numerator

0

47

48

49

49

49

49

Denominator

49

49

49

49

49

49

49

Percent Growth

0

95.9

98.0

100

100

100

100