FEATURED PRODUCT: Interpretation of FDA's (QSR) With QSIT references
Search This Site
Biological Indicator Reduced Incubation Time Study
(RIT)
Medical Device ethylene oxide sterilization is
monitored by the use of biological indicators (BIs) using Bacillus subtilis
species. The common “off the shelf” recommendation for incubation by most
manufacturers is 7 days or more. However, there is a way to reduce the
incubation time to as much as 24 hours.
Reducing the BI incubation time can be a big financial asset to an organization
bent on reducing inventory. Other benefits including being able to matching
customer demand and just having better overall control with regards to timing.
A lot of factors take part in the RIT including the sterilization cycle and the
Challenge Pack (developed to mimic the device packaging – either equaling or
making it more difficult for gas penetration). When a manufacturer decides to
reduce the BI incubation time the method chosen to validate the process is very
crucial in ensuring a) a successful time reduction and b) ensuring that the
process is adequately validated to meet FDA requirements.
FDA Requirements
The FDA has proposed a method for validating a RIT. In summary the process works
by subjecting a Lot of BIs to a Fractional cycle (or sublethal cycle – since the
exposure time is significantly reduced). If done correctly a number of BIs will
grow out. It is critical that the number stay within a range. The FDA recommends
30 to 80 percent survival.
The following is an excerpt taken from Premarket Notifications [510(k)] for
Biological Indicators Intended to Monitor Sterilizers Used in Health Care
Facilities; Draft Guidance for Industry and FDA Reviewers
You can read the document online at:
http://www.fda.gov/cdrh/ode/guidance/1320.html#tablenote_1
Obtain a minimum of 300 biological indicators. One hundred (100) biological
indicators should be from each of 3 separate lots.
Using the predefined sterilization parameters, expose the biological indicators
in 3 partial sterilization cycles (100 per partial cycle). Each of the test
cycles will have 30% to 80% of the indicators surviving (i.e., test positive).
A partial cycle is one in which all sterilization parameters, except the time
parameter, are met. The exposure time is much shorter than the standard
sterilization cycle.
Only the biological indicators from one lot are to be used in each partial
cycle. Do not mix biological indicators from different lots.
It is preferable to run biological indicators in a device load. However, the
inherent difficulties of achieving a partial cycle kill under such circumstances
are understood. Thus, partial cycles can be run without device loads.
NOTE: During all sterilization validation studies, consider the effects of the
sterilant in combination with the device material on the indicator organism. If
the materials are judged to have a significant effect on organism destruction,
the biological indicators should be exposed to the sterilant in conjunction with
the devices during the partial cycle studies.
If there are fewer than 30% survivors or more than 80% survivors in any one run,
this particular run is invalid and should be rerun to achieve the desired number
of survivors.
Three partial cycles are the minimum number of testing cycles to be run. If the
results of any one cycle are invalid, another partial cycle should be
substituted for it.
After exposure incubate the BIs for a minimum of 7 days. Place the BIs in the
growth media no more than 8 hours after removal from the sterilization chamber
or removal from the sterilized load of devices. Record the number of positive
BIs on either a daily basis or for the particular time interval of interest.
Using the number of Bls that test positive on day 7 as the base of 100% grow out
(denominator data), determine from the growth chart if the required number of
BIs have grown out (numerator data) in the time interval in question. More than
97% of the base number of BIs should test positive in each partial cycle for the
proposed incubation time to be acceptable.
The greatest number of days of incubation required to obtain more than 97%
positive BIs (based on the 7 day incubation time) in any one of the partial
cycles is the minimum incubation time that will be allowed. Averaging the three
(or more) partial cycle incubation times is not allowable (see example in
Appendix).
If the biological indicator user has not or cannot validate the biological
indicator incubation period using the described methodology, then the user
should remain with at least a seven day incubation period.
Number Positive Biological Indicators Required to Achieve 97% Growth
|
Data1 |
30 |
31 |
32 |
33 |
33 |
34 |
35 |
36 |
37 |
38 |
|
|
Denominator Data2 |
30 |
31 |
32 |
33 |
34 |
35 |
36 |
37 |
38 |
39 |
|
|
Numerator Data |
39 |
40 |
41 |
42 |
43 |
44 |
45 |
46 |
47 |
48 |
|
|
Denominator Data |
40 |
41 |
42 |
43 |
44 |
45 |
46 |
47 |
48 |
49 |
|
|
Numerator Data |
49 |
50 |
51 |
52 |
53 |
54 |
55 |
56 |
57 |
58 |
|
|
Denominator Data |
50 |
51 |
52 |
53 |
54 |
55 |
56 |
57 |
58 |
59 |
|
|
Numerator Data |
59 |
60 |
61 |
62 |
63 |
64 |
65 |
65 |
66 |
67 |
|
|
Denominator Data |
60 |
61 |
62 |
63 |
64 |
65 |
66 |
67 |
68 |
69 |
|
|
Numerator Data |
68 |
69 |
70 |
71 |
72 |
73 |
74 |
75 |
76 |
77 |
78 |
|
Denominator Data |
70 |
71 |
72 |
73 |
74 |
75 |
76 |
77 |
78 |
79 |
80 |
The numerator is the number of positive biological indicators that is greater
than 97% of the denominator. If the numerator is equal to or greater than the
one listed for the corresponding denominator (based on the total number of
positive biological indicators on day 7), the length of the incubation time when
this occurs is acceptable.
2 The denominator is the total number of positive biological indicators on day 7
of incubation.
Example: A manufacturer would like to reduce its biological indicator incubation
time to 3 days. The test data show the following:
|
|
Day 1 |
Day 2 |
Day 3 |
Day 4 |
Day 5 |
Day 6 |
Day 7 |
|
Partial cycle #1 |
|
|
|
|
|
|
|
|
Numerator |
0 |
56 |
57 |
57 |
58 |
59 |
59 |
|
Denominator |
59 |
59 |
59 |
59 |
59 |
59 |
59 |
|
Percent Growth |
- |
94.9 |
96.6 |
96.6 |
98.3 |
100 |
100 |
|
|
|
|
|
|
|
|
|
|
Partial cycle #2 |
1 |
34 |
35 |
35 |
35 |
35 |
35 |
|
Numerator |
35 |
35 |
35 |
35 |
35 |
35 |
35 |
|
Denominator |
2.9 |
97.1 |
100 |
100 |
100 |
100 |
100 |
|
Percent Growth |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Partial cycle #3 |
|
|
|
|
|
|
|
|
Numerator |
0 |
79 |
81 |
Test invalid because percent positive biological indicators is outside the allowable window (30-80% |
|||
|
Denominator |
|
|
|
||||
|
Percent Growth |
|
|
|
||||
|
|
|
|
|
|
|
|
|
|
Partial cycle #4 |
|
|
|
|
|
|
|
|
Numerator |
0 |
47 |
48 |
49 |
49 |
49 |
49 |
|
Denominator |
49 |
49 |
49 |
49 |
49 |
49 |
49 |
|
Percent Growth |
0 |
95.9 |
98.0 |
100 |
100 |
100 |
100 |
The 3-day reduced incubation time is not validated because, of the three valid
partial cycles, not all achieved 97% growth in 3 or fewer days. However, based
upon the criteria listed in #5 of the test methodology, the data validate a
5-day reduced incubation time. The 5-day incubation time in this example is the
greatest number of days, from all the valid partial cycles, needed to grow out
more than 97% of the denominator biological indicators.
* * * * * * * * * * * * * * * * *
Bier Vessels
The best way to conduct a RIT study is using a Bier vessel in a laboratory
setting without the device load. Though many have argued that a Bier vessel does
not closely enough mimic the “real” conditions there are many benefits.
For those of you not familiar, a bier vessel is a miniature sterilization
chamber. The vessel allows for the rapid evacuation of gases, the rapid
injection of Ethylene Oxide gas and finally the subsequent evacuation and
elimination of sterilant from the chamber. This last point is very important for
a successful RIT study. It is known that ‘residual kill’ can occur in the post
gas dwell phase. The time to eliminate the gas is critical in ensuring
repeatable results. Likewise using a full sterilization chamber can have
unpredictable results – often costing you in a lot of unsuccessful runs.
Conclusion
In conclusion an ethylene oxide sterilization program can in fact compete with
the quick turn-times seen by Gamma sterilization. Unless extra inventory and
turn-time has no real benefits to your organization a RIT study can in turn
quickly pay for itself. A reason RIT may not benefit you is if your product
requires a lengthy degassing (or aeration) phase. Some products require weeks if
not months before levels of Ethylene Oxide residuals drop to acceptable levels.
If residual levels are not a concern having a RIT study as stated above can be
rewarding.
Site Map »
Media package available for advertisers looking to advertise on this site.
A
complete compendium to FDA's Quality System Regulation (QSR) with relative
references to FDA's Quality System Inspection Technique (QSIT). Now ONLY $99 (limited time)